CVE-2021-32993
THREAT INTELLIGENCE REPORT — CVE-2021-32993
Executive Summary
CVE-2021-32993 is a critical hard-coded credential vulnerability affecting the Philips IntelliBridge EC 40 and EC 60 Hub (firmware C.00.04 and prior). These devices act as integration gateways between patient monitoring equipment and hospital IT networks. The vulnerability allows an attacker with network access — or physical access in some deployment scenarios — to obtain and use embedded credentials (passwords or cryptographic keys) that the device uses for inbound authentication, outbound communication to external components, or internal data encryption. With a CVSS score of 8.1, successful exploitation grants an attacker the ability to authenticate as the device itself, potentially intercepting, modifying, or injecting data flowing through the hub, compromising the confidentiality and integrity of patient telemetry data. This is a supply-chain-level defect introduced by the vendor and remains unpatched.
This record — and 170,000+ others — are available via the ValtersIT API. One endpoint replaces five: NVD + GitHub CVE + CISA KEV + EPSS + Nuclei + Exploit-DB, unified.
No signup required
Technical Analysis
The IntelliBridge EC 40 and EC 60 are embedded network hubs that bridge bedside patient monitors (e.g., Philips IntelliVue monitors) to hospital information systems (HIS) and electronic medical records (EMR). They contain firmware version C.00.04 and earlier versions, which embed hard-coded credentials directly in the binary image (flash memory).
Vulnerability Mechanism
Per the NVD description, the device "contains hard-coded credentials, such as a password or a cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data." This is a classic CWE-798 flaw introduced at build time.
Three distinct credential usage patterns are implicated:
Inbound Authentication — The device has an embedded administrative or service credential that allows local/remote login (e.g., via SSH, HTTPS management interface, or a diagnostic serial port). These credentials are identical across all devices worldwide and cannot be changed by the operator (or are reset to defaults on firmware upgrade).
Outbound Communication — The hub authenticates to upstream systems (e.g., EMR middlewares, integration engines, or Philips-specific backend services) using a fixed credential. An attacker who extracts this credential can impersonate the hub, submitting false patient data to the EMR, or sniffing legitimate data to identify the credential in transit.
Internal Data Encryption — If the credential is used as a cryptographic key (or as the seed to derive one), an attacker in possession of that credential can decrypt any data at rest on the device (e.g., cached telemetry, configuration logs) or decrypt network traffic that the device encrypts.
The hard-coded nature means that no per-device randomization exists — the credential can be recovered from one unit and applied against all other units of the same model/firmware. The credential is recoverable via binary reverse engineering of a downloaded firmware image or via a serial/device debug interface.
Attack Surface
- Network: The hub is connected to the hospital LAN (often VLAN-isolated but not always). If the management interface or communication endpoints are reachable remotely, an attacker can attempt authentication using the known credential.
- Physical: An attacker with brief physical access (e.g., a cleaning staff member, a visitor, or a contractor) can connect a laptop to the device's serial port or Ethernet management port and attempt login with the embedded credential.
- Supply Chain: A malicious actor in the firmware distribution chain (or a reseller) can extract the credential from a single firmware image and then build weaponized tooling.
Code-Level Explanation (Inferred)
The NVD description does not provide specific function names, file paths, or API endpoints. That said, the typical implementation pattern for such embedded systems is:
- Credentials stored in a plaintext configuration file (e.g., a
.conffile) or in a hard-coded string table within the firmware executable. - Common locations include
/etc/passwd,/etc/shadow(or equivalent), or string constants in the main binary. - The device software compares user-supplied credentials against these baked-in values without a mechanism to override them at runtime.
No specific code identifiers are available in the CVE data; the attack surface description above is based on the general behavior of the device class.
Exploitation Scenarios
Prerequisites for Exploitation
- Network access to the management interface of the IntelliBridge hub, OR physical access to the device's debug ports, OR access to one of the outbound communication endpoints (e.g., an EMR integration engine) that the hub authenticates against.
- The attacker must possess the hard-coded credential. This can be obtained by:
- Downloading a firmware update image from Philips (or a third-party mirror) and extracting the credential via firmware analysis tools (e.g., binwalk, strings).
- Physical extraction from a unit acquired via secondary market or e-waste.
- Public vulnerability disclosures or leak sites.
Attack Chain (Inferred from Vulnerability Class)
Reconnaissance / Target Identification: The attacker scans the hospital network (or the internet if the device is exposed) for TCP ports commonly used by the device — typically 80/443 (web UI), 22 (SSH), 23 (telnet), or the proprietary integration port. Shodan dorks such as
title:IntelliBridgeorPhilips IntelliBridgecan identify exposed instances. Once a target is found, the attacker determines the firmware version (e.g., "C.00.04") by probing the web UI or banner grabbing.Flaw Trigger: Using the recovered hard-coded credential (e.g., a default admin password like
admin:adminor a service account likeintellibridge:bridgeBridge123— hypothetical examples, not confirmed), the attacker authenticates to the device's management interface or directly to the outbound integration service. The device accepts this credential because it matches its embedded static value.Payload / Execution: At the vulnerability-class level, the attacker:
- Gains authenticated remote access to the hub's management shell/API.
- Can modify device configuration (e.g., change where data is forwarded, disable logging, or install a reverse proxy).
- Can inject data into the outbound telemetry stream (emulating a patient monitor) or extract/observe the live monitoring data.
- If the embedded credential is the encryption key, the attacker can passively decrypt captured traffic.
Post-Exploitation Impact: The attacker establishes persistent, stealthy access to patient data. They can:
- Trigger false alarms/silent alarms (affecting patient safety).
- Corrupt the EMR, causing misdiagnosis.
- Use the hub as a pivot point to move laterally within the hospital network.
- Exfiltrate PHI, leading to HIPAA violation fines and reputation damage.
Attack Feasibility Notes
- No public PoC or exploit code is available at the time of writing. The attack chain is inferred from the vulnerability class, not from a demonstrated exploit.
- The CVSS subscores (AV:N, AC:H, PR:N, UI:N, S:U, C:H/I:H/A:H) indicate that while network-exploitable, the attack complexity is HIGH, likely due to the need to obtain the credential (a non-trivial reverse-engineering step).
Impact Assessment
Business Impact
- Confidentiality: High (CVSS C:H) — Live patient vital signs (heart rate, blood pressure, SpO2, etc.) are transmitted through the hub. An attacker can capture this data, violating HIPAA/patient privacy regulations.
- Integrity: High (CVSS I:H) — An attacker can inject falsified patient data into the EMR, which could lead to incorrect clinical decisions with life-threatening consequences.
- Availability: High (CVSS A:H) — The attacker can disrupt device connectivity, causing monitor disconnect alarms, or brick the device by changing configuration/firmware.
Blast Radius — The affected devices are used extensively in hospitals, ICUs, and surgical wards. A single compromised hub can affect all patients connected to it. In many hospital network deployments, the hubs are not segmented from the rest of the clinical network, so a compromised hub could be used to reach other clinical systems (imaging, labs, pharmacy).
Relevance — Philips has issued advisories for the IntelliBridge series before (e.g., other CVEs on similar devices). The medical device sector is under increasing scrutiny by regulators (FDA) and attackers (e.g., ransomware groups targeting hospitals). While CISA KEV does NOT list this CVE, the EPSS score of 0.00272 (19.4 percentile) indicates low immediate exploit likelihood but nonzero.
Detection Guidance
Sigma Rule
title: Philips IntelliBridge Hub Authentication Attempt with Known Hardcoded Credentials
id: 74f3bcd0-3e9e-4c2f-b5af-81a6f5e1b3a2
status: experimental
description: Detect login attempts to Philips IntelliBridge EC 40/60 Hub using credentials matching known hardcoded defaults
author: vartersit SOC
logsource:
product: suricata
category: http
detection:
selection:
http.url:
- '*/login*'
- '*/auth*'
- '*/main*'
http.http_user_agent: 'PACX*'
filter:
http.response_code: 200
condition: selection and not filter
level: high
falsepositives:
- Legitimate vendor maintenance with factory-restored devices
- Integration testing during rollout
fields:
- src_ip
- dest_ip
- http.url
- http.http_user_agent
tags:
- attack.t1190
- cve.2021-32993
- cwe.798
YARA Rule
rule Philips_IntelliBridge_Hardcoded_Creds_CVE2021_32993 {
meta:
author = "vartersit.com CTI"
description = "Detects firmware and files containing hardcoded credentials used by Philips IntelliBridge EC 40/60"
cve = "CVE-2021-32993"
date = "2024-01-15"
severity = "high"
strings:
$cred_charset = /[A-Za-z0-9_\-!@#$%^&*+]{8,32}/
$fw_string = "IntelliBridge" ascii nocase
$version = "C." ascii
condition:
$fw_string and $version and any of them
}
Suricata IDS Rule
alert tcp any any -> any 80 (msg:"PHILIPS IntelliBridge Hardcoded Credential Login Attempt (CVE-2021-32993)"; flow:to_server,established; content:"POST"; http_method; content:"login|in|php"; http_uri; content:"user="; nocase; http_client_body; content:"pass="; nocase; http_client_body; threshold:type both, track by_src, count 3, seconds 60; reference:cve,2021-32993; classtype:attempted-admin; sid:1005293; rev:1;)
Elastic EQL Rule (Elastic Detection Rules)
sequence by host.ip with maxspan=30s
[network where network.protocol=="http" and destination.ip == "10.x.x.x" and http.request.method=="POST"]
[authentication where event.outcome=="success" and winlog.event_id==4624]
A more specific rule for the device logs (if available):
any where
event.dataset == "intellibridge" and
event.action in ("login","auth_fail") and
user.name in ("admin","service","root")
Splunk SPL Query
index=main sourcetype=intellibridge*
| search "login" OR "auth" OR "admin"
| rex field=message "user=(?<uname>\w+)"
| where uname IN ("admin","service","root")
| stats count by src_ip, src_port, uname, _time
| table _time, src_ip, src_port, uname, count
Wazuh OSSEC Rule
<group name="philips-intellibridge,">
<rule id="100250" level="10">
<decoded_as>json</decoded_as>
<field name="device">IntelliBridge</field>
<match>hardcoded|intellibridge|default_credential</match>
<description>Philips IntelliBridge Hardcoded Credential Attempt/Use Detected</description>
<options>no_full_log</options>
<group>authentication_failures,</group>
</rule>
</group>
Mitigation & Patch Guidance
Patch Status: UNPATCHED (0/4 confidence). No vendor advisory, security update, or firmware fix has been identified from any tracked source. The affected firmware version is C.00.04 and all prior versions. Philips has not released a public remediation.
Immediate Mitigations (compensating controls)
- Network Segmentation — Isolate all IntelliBridge hubs in a dedicated VLAN with strict egress/ingress ACLs. Only allow traffic to the EMR integration endpoints, and block all remote management access (disable web UI, SSH, telnet) unless from an admin jump host.
- Disable Unused Services — If the management UI is not required for daily operations, turn it off at the network level (block ports 80/443/22/23 to the device).
- Change Device Configuration — Check the device configuration to see if credentials are configurable. If the device supports operator-defined credentials, change them immediately. If not, treat the device as untrusted.
- Monitor Traffic — Deploy the detection rules from this report to identify any login attempts to the hub using default/admin credentials. Alert on any successful authentication events from non-admin IPs.
- Physical Security — Lock racks and access to the device serial ports.
Patch / Upgrade Steps
- None available. However, monitor Philips Security Advisory mailing list and ICS-CERT/MedCyber advisories for a firmware update (C.00.05 or newer) that replaces static credentials with per-device generated keys.
- If a newer firmware becomes available, validate it on a test bench with updated detection rules before deployment.
Long-Term Hardening
- Asset Inventory — Maintain a CMDB entry for every IntelliBridge hub (IP, firmware, location) to facilitate patching when a fix is released.
- Vendor Risk Management — Add a contractual requirement for Philips to disclose and remediate hardcoded credential findings. Consider this vulnerability in future procurement decisions.
- Incident Response Plan — Prepare a playbook for hub compromise: quarantine steps, forensic imaging, PHI breach notification, and backup EMR data integrity checks.
- Credential Hygiene — Since these are OT/devices, perform regular firmware checksum validation to detect tampering.
References
- NVD Entry — https://nvd.nist.gov/vuln/detail/CVE-2021-32993
- Philips Product Security Advisory — check https://www.philips.com/a-w/security/security-advisories.html for future updates (no advisement issued as of report date).
- CISA ICS-CERT Advisories (related devices) — https://www.cisa.gov/ics/advisories (search "IntelliBridge for related disclosures).
- MITRE ATT&CK T1587.001 — https://attack.mitre.org/techniques/T1587/001/
- EPSS Data (FIRST) — https://www.first.org/epss/ (score 0.00272, 19.4% percentile).
No threat-actor activity or in-the-wild exploitation has been publicly documented for this CVE. The CISA KEV database does not include this CVE, and no exploit code has been released. Risk is assessed as moderate due to the medical context and high CVSS base score, but practical exploit complexity is high given the requirement to retrieve the hard-coded credential via firmware reverse engineering.
End of Report — Prepared by vartersit.com CTI Team.