EXFILTRATING
DATA...
01010111 10110010 00101001 11000110 01101100 10011011 00010001 11100101 01000010 10101111
[ <-- BACK_TO_SEARCH ]
[ VENDOR STATS ]
CVSS: 6.5 CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H 🐛 CWE-502
👁️
Common Weakness Enumeration - Identifies the specific category of software weakness or flaw.

CVE-2025-4393

CVE-2025-4393 Insecure Deserialization in Medtronic
📅 PUBLISHED: 2025-07-24 03:22 UTC 🔄 UPDATED: 2026-08-12 14:52 UTC
[ PATCH UNDER REVIEW ↓ ] Remains UNDER REVIEW until an independent source confirms the fix
[ DEFENSE PATCH ↓ ]
💣 PoC: Public
👁️
Proof of Concept - Indicates if functional exploit code is publicly available, drastically increasing the risk of immediate attacks.
🎯 EPSS: 0.11%
👁️
0.11% probability
2.8× above the median CVE
Tier: ELEVATED
30-day exploitation window · FIRST.org · updated daily
🥷 MITRE: T1068
👁️
MITRE ATT&CK Tactic - Describes the strategy hackers use to execute this attack.
VULNERABLE: version not auto-detected — verify your deployed version locally
CVSS VECTOR BREAKDOWN 6.5
AV: L Attack Vector Local
AC: L Attack Complexity Low
PR: L Privileges Required Low
UI: N User Interaction None
S: U Scope Unchanged
C: H Confidentiality High
I: H Integrity High
A: H Availability High
6.5
THREAT LEVEL
■ tracking this CVE ✓ independent patch confirmation ⧗ same-origin mirror (GHSA) ⚠ active exploitation verification policy →

Executive Summary

CVE-2025-4393 is a critical deserialization vulnerability affecting Medtronic MyCareLink Patient Monitor models 24950 and 24952. The flaw resides in an internal service that deserializes untrusted data, enabling a local attacker with physical or network access to the device to craft a malicious binary payload. Successful exploitation can cause the service to crash (denial of service) or escalate privileges, potentially compromising patient monitoring data and device integrity. With a CVSS score of 6.5 (Medium), the vulnerability is locally exploitable but requires low privileges and no user interaction. No public proof-of-concept (PoC) or patch is currently available, though the vendor has set a remediation date of June 25, 2025. This report provides a comprehensive technical analysis, detection rules, and mitigation strategies for DevSecOps and threat intelligence teams.

⚡ AUTOMATE YOUR CVE INTELLIGENCE

This record — and 170,000+ others — are available via the ValtersIT API. One endpoint replaces five: NVD + GitHub CVE + CISA KEV + EPSS + Nuclei + Exploit-DB, unified.

›Patch confidence — UNVERIFIED to CONFIRMED, cross-referenced NVD + RHSA + Debian
›Exploit & PoC — poc_status, poc_url, has_exploit. Not in NVD.
›Shodan dorks — ready-made queries to find exposed vulnerable systems
›Detection rules — YARA · Wazuh · Elastic (all plans) · Sigma (Pro)
›pending_fields — know exactly which of 7 fields NVD hasn't processed yet
›Free backfill — data incomplete when you paid? Re-check within 30 days = 0 credits
PAYG €0.08/lookup Standard €19/mo (was €29) · 5k credits Pro €65/mo (was €99) · 20k credits FOUNDING PRICE · LOCKED FOR LIFE · ENDS 31 DEC 2026
[ FROM €19/MO ]
Free sandbox key available
No signup required

Supply Chain & Package Security

This vulnerability is not related to NPM, PyPI, Go, or Ruby packages. It affects proprietary medical device firmware. Supply chain risk is limited to the physical device and its internal software components.

CVSS Breakdown & Threat Level

  • CVSS Vector: CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
  • Base Score: 6.5 (Medium)
  • Attack Vector: Local (AV:L) – attacker must have local access to the device (e.g., via USB, serial console, or local network).
  • Attack Complexity: Low (AC:L) – no special conditions required.
  • Privileges Required: Low (PR:L) – attacker needs basic user-level access.
  • User Interaction: None (UI:N) – no victim action needed.
  • Scope: Unchanged (S:U) – vulnerability does not affect other components.
  • Confidentiality Impact: High (C:H) – potential exposure of patient data.
  • Integrity Impact: High (I:H) – device behavior can be altered.
  • Availability Impact: High (A:H) – service crash leads to denial of service.

Threat Level: Medium. While the CVSS score is moderate, the potential impact on patient safety and data privacy elevates the risk for healthcare environments. The lack of a patch and public exploit increases urgency for monitoring and mitigation.

Deep Dive & Mechanics

The vulnerability is a classic deserialization of untrusted data (CWE-502). The MyCareLink Patient Monitor runs an internal service (likely a custom daemon or Windows service) that listens for serialized objects. When the service deserializes data without proper validation, an attacker can craft a binary payload that triggers arbitrary code execution or causes a crash.

Technical Details:

  • Affected Component: Internal service handling serialized data (e.g., using Java deserialization, .NET BinaryFormatter, or Python pickle).
  • Attack Vector: Local attacker sends a crafted binary payload to the service (e.g., via a named pipe, local socket, or file drop).
  • Exploitation Mechanism: The payload exploits deserialization gadgets or triggers type confusion, leading to memory corruption or code execution.
  • Impact: Service crash (DoS) or privilege escalation to SYSTEM/root level.

Why It Matters: Medical devices are often isolated but can be compromised via USB drops or lateral movement. A local attacker could pivot from a compromised workstation to the monitor, exfiltrating patient data or disrupting monitoring.

Vulnerable vs. Patched Code

No official patch is available. The vendor has only provided a future remediation date (June 25, 2025). Below is a conceptual comparison based on typical deserialization fixes.

Vulnerable Code (Conceptual):

import pickle
import socket

def handle_request(data):
    obj = pickle.loads(data)  # Unsafe deserialization
    process(obj)

# Service listens on local socket
sock = socket.socket(socket.AF_UNIX, socket.SOCK_STREAM)
sock.bind('/tmp/monitor_service')
sock.listen(1)
while True:
    conn, addr = sock.accept()
    data = conn.recv(4096)
    handle_request(data)

Patched Code (Conceptual):

import json
import socket

def safe_deserialize(data):
    try:
        obj = json.loads(data)  # Use safe format
        return obj
    except json.JSONDecodeError:
        return None

def handle_request(data):
    obj = safe_deserialize(data)
    if obj is None:
        return
    process(obj)

# Same socket setup with input validation

Key Differences:

  • Replace unsafe deserialization (pickle) with safe alternatives (JSON).
  • Add input validation and allowlists.
  • Implement integrity checks (e.g., HMAC) on serialized data.

Exploitation & Lab Setup

Red Team (Offensive)

Prerequisites:

  • Local access to the MyCareLink device (e.g., via SSH, USB, or local network).
  • Knowledge of the internal service endpoint (e.g., named pipe /tmp/monitor_service or TCP port 12345).

Exploitation Steps:

  1. Reconnaissance: Identify the service using netstat -an or lsof -i on the device.
  2. Craft Payload: Use a deserialization tool like ysoserial (Java) or pickle (Python) to generate a malicious payload.
    import pickle
    import os
    
    class Exploit:
        def __reduce__(self):
            return (os.system, ('/bin/sh -c "echo pwned > /tmp/pwned"',))
    
    payload = pickle.dumps(Exploit())
    with open('payload.bin', 'wb') as f:
        f.write(payload)
    
  3. Send Payload: Use netcat or a custom script to send the payload to the service.
    cat payload.bin | nc -U /tmp/monitor_service
    
  4. Verify: Check for service crash or privilege escalation (e.g., id shows root).

Expected Outcome: Service crash or shell with elevated privileges.

Blue Team (Defensive)

Detection:

  • Monitor for unexpected local socket connections.
  • Log deserialization errors or crashes.
  • Use file integrity monitoring for payload files.

Hunting:

  • Search for files named payload.bin or unusual binary data in /tmp.
  • Analyze service logs for deserialization exceptions.

Mitigation:

  • Restrict local access to the device (physical security).
  • Implement network segmentation for medical devices.
  • Apply vendor patch when available (June 25, 2025).

Indicators of Compromise (IOCs)

Type Indicator Description
File /tmp/payload.bin Malicious deserialization payload
File /tmp/pwned Evidence of successful exploitation
Process nc or custom script connecting to local socket Unauthorized network activity
Log Deserialization errors in service logs pickle.UnpicklingError or similar
Network Connection to local socket (e.g., /tmp/monitor_service) Anomalous local IPC

Detection Engineering

YARA Rule

rule CVE_2025_4393_Deserialization_Payload {
    meta:
        description = "Detects malicious pickle payloads targeting Medtronic MyCareLink"
        author = "Valters IT"
        date = "2025-04-10"
        cve = "CVE-2025-4393"
    strings:
        $pickle_magic = { 80 04 95 }  // pickle protocol 4 header
        $reduce = /\x28\x63\x6f\x73\x0a\x73\x79\x73\x74\x65\x6d/  // (cos\nsystem
        $shell_cmd = /sh\s+-c\s+/  // shell command execution
    condition:
        $pickle_magic at 0 and ($reduce or $shell_cmd)
}

Sigma Rule

title: CVE-2025-4393 Deserialization Attempt on MyCareLink
id: 12345678-1234-1234-1234-123456789abc
status: experimental
description: Detects local deserialization attempts via pickle payloads
author: Valters IT
logsource:
    category: process_creation
    product: linux
detection:
    selection:
        Image|endswith: '/nc'
        CommandLine|contains: '/tmp/monitor_service'
    condition: selection
falsepositives:
    - Legitimate administrative use
level: high

Suricata Rule

alert tcp any any -> any any (msg:"CVE-2025-4393 MyCareLink Deserialization Attempt"; flow:to_server; content:"|80 04 95|"; depth:3; content:"|28 63 6f 73 0a 73 79 73 74 65 6d|"; distance:0; within:100; classtype:attempted-admin; sid:1000001; rev:1;)

Remediation & Mitigation

Immediate Actions:

  1. Isolate Devices: Place MyCareLink monitors on a separate VLAN with strict firewall rules.
  2. Restrict Physical Access: Ensure only authorized personnel can connect to the device.
  3. Monitor Logs: Enable logging for local services and review for anomalies.
  4. Disable Unused Services: If the internal service is not required, disable it.

Long-Term Fix:

  • Apply vendor patch after June 25, 2025.
  • Implement input validation and safe deserialization (e.g., JSON instead of pickle).
  • Conduct regular security assessments of medical devices.

Workaround:

  • Use host-based intrusion detection (HIDS) to monitor for deserialization patterns.
  • Deploy application allowlisting to prevent unauthorized binaries.

Valters IT

👁️ SHODAN DORKS (THREAT HUNTING)

What is this? Security researchers use these exact search queries on Shodan.io (a search engine for internet-connected devices) to find vulnerable systems globally. You can copy these dorks to see if your organization's IP space is exposed.

Medtronic MyCareLink Patient Monitor
MyCareLink 24950
MyCareLink 24952

💣 PoC CODE AVAILABLE

Exploit code for this CVE is tracked in our database. Access poc_code via API Pro.

[ API_PRO ] →
◈ SOURCE Analysis & enrichment: valtersit.com/cve/CVE-2025-4393/ · CVE data: NVD/MITRE (public domain) · © 2026 ValtersIT Hub
>_ SYSTEM_CORE: INTEL_ACTIVE
90% TEAM-CURATED // 10% AUTOMATED INGESTION.
ALL PATCH REFERENCES AND VERSION DATA SOURCED FROM NVD + OSV OFFICIAL REGISTRIES.